Original Article | Volume 16 Issue 11 (November, 2026) | Pages 24 - 31
Arrhythmia detection by Extended ambulatory continuous electrocardiography patch monitoring system in a tertiary care hospital in India
 ,
 ,
 ,
 ,
 ,
 ,
 ,
1
Department of Cardiology, PSG Hospitals, Peelamedu, Coimbatore, Tamil Nadu 641004, India.
Under a Creative Commons license
Open Access
Received
Sept. 24, 2026
Revised
Oct. 5, 2026
Accepted
Oct. 21, 2026
Published
Nov. 6, 2026
Abstract

Background: Ambulatory electrocardiography is routinely used to detect paroxysmal arrhythmias. Although 24-hour Holter monitoring is commonly used, its diagnostic yield for infrequent arrhythmias is limited. WebCardio is a patch-based ambulatory monitoring device that can record for 3 days or up to 7 days and may improve arrhythmia detection. Objective: To evaluate the utility of Web Cardio, a patch-based device that uses two-channel ECG monitoring with cloud-based storage of recorded rhythms for 3 days or 7 days, in patients with symptoms suggestive of paroxysmal arrhythmias. Methods: The Web Cardio patch device was attached to patients with suspected paroxysmal arrhythmias in the Department of Cardiology, PSGIMSR, for 3 days (3-day cohort) or for up to 7 days (7-day cohort), and the recordings were evaluated. The two cohorts were analyzed separately and were not pooled. The ability to detect any of the following arrhythmias was analyzed: atrial fibrillation (AF), atrioventricular (AV) block, sinus pause of more than 2.5 seconds, supraventricular tachycardia (SVT), ventricular tachycardia (VT). Results: In the 3-day cohort, 1076 patients with symptoms suggestive of paroxysmal arrhythmias underwent Web Cardio patch monitoring (mean age 55.86 ± 19.60 years; 639 [59.39%] male and 437 [40.61%] female). A total of 1076 patients were monitored, the most frequent being supraventricular tachycardia (SVT; 472), ventricular tachycardia (VT; 191), second-degree AV block (109), pauses (100), atrial fibrillation (AF; 89), complete heart block (CHB; 37), and >2:1 AV block (4). A substantial proportion of events were first detected after the initial 24 hours: 30.95% of AF, 35.17% of SVT, 40.84% of pauses, and 44.0% of VT. In the separate 7-day cohort (n = 29; mean age 66.41 ± 14.71 years; 16 male [55.17%] and 13 females [44.83%]), at least one arrhythmia was detected in 25 patients (86.21%): SVT in 19 (65.52%), AV block in 9 (31.03%), pauses in 9 (31.03%), VT in 4 (13.79%), and AF in 1 (3.45%). Findings with potential management relevance, including conduction disease, pauses, VT, and AF, were identified in both cohorts and are relevant to decisions on pacing, anticoagulation, and further electrophysiological evaluation. Conclusion: Prolonged, continuous patch-based ECG monitoring with the Web Cardio system demonstrated clinical utility by detecting a substantial burden of clinically actionable arrhythmias, including significant conduction abnormalities, pauses, ventricular arrhythmias, and atrial fibrillation. In the 3-day cohort (n = 1,076), a considerable proportion of AF, SVT, pauses, and VT events were first detected after 24 hours, suggesting that shorter recordings may under-detect clinically relevant arrhythmias. In the separate 7-day cohort (n = 29), at least one arrhythmia was detected in 25 patients (86.21%), most commonly SVT (65.52%), followed by AV block and pauses (31.03% each) and VT (13.79%). The cohorts were analyzed separately and were not pooled. The detection of conduction disease, pauses, VT, and AF provides clinically meaningful information with potential implications for pacing evaluation, anticoagulation, antiarrhythmic therapy, and electrophysiological referral. Although downstream management decisions and patient outcomes were not assessed, these findings support the diagnostic and potential clinical utility of extended WebCardio patch monitoring in patients with suspected paroxysmal arrhythmias.

Keywords
INTRODUCTION

The prevalence of arrhythmias is increasing as life expectancy improves with advances in healthcare. A 12-lead electrocardiogram is usually the first investigation for patients presenting with paroxysmal palpitations or syncope. However, a 12-lead ECG often does not capture the diagnosis. Although 24-hour Holter monitoring has been routinely used for suspected paroxysmal arrhythmias, it can be cumbersome, and its sensitivity for detecting intermittent arrhythmias is limited. Several ambulatory ECG devices are available, ranging from event monitors to implantable loop recorders; however, there is no standard duration or device for evaluating paroxysmal arrhythmias, and there is no consensus on the ideal device or monitoring period. Therefore, we conducted a prospective study at our institute using a 3-day (and, in a separate cohort, 7-day) externally applied WebCardio patch monitor in patients with suspected paroxysmal arrhythmias to evaluate the safety, efficacy, and diagnostic yield of arrhythmia detection.

 

Objective:

To assess the utility of a web-based external patch monitor for detecting arrhythmia burden in patients with suspected paroxysmal arrhythmias.

MATERIALS AND METHODS

In all patients, the Webcardio patch was applied to the left upper chest after skin preparation. The Webcardio patch is a single-use, two-channel ECG patch that weighs approximately 28 grams and measures 10 cm × 8 cm. It can be applied easily to the chest wall and is comfortable for patients. The patch was applied after obtaining consent, and patients were instructed on how to trigger the device during symptoms. The Webcardio patch was paired with a mobile device containing the WebCardio app, downloaded from the Google Play Store or Apple store. Relevant patient details were entered into the WebCardio app at the start of the procedure, and patients were instructed to trigger the device during symptoms. After the initial setup, ECG signal quality was confirmed using the WebCardio system. Once paired, the Webcardio patch continuously and wirelessly transmitted ECG signals to the mobile device and then to the remote server until disconnection or up to 3 days (3-day cohort) or up to 7 days (7-day cohort). All patients were encouraged to keep the mobile device with them throughout monitoring. The ambulatory ECG was analysed using proprietary software by a qualified Holter technician at the remote centre, and a report was generated and sent to the hospital by email. The investigating cardiologists reviewed the complete ambulatory ECG recording, arrhythmic events, and report after logging in to the WebCardio site. Detection of at least one of the following arrhythmias was assessed: atrial fibrillation (AF), atrioventricular (AV) block, sinus pause of more than 2.5 seconds, supraventricular tachycardia (SVT), premature ventricular complexes (PVCs), ventricular tachycardia (VT), and ventricular fibrillation (VF). Standard definitions of arrhythmia were used for event review.

 

In addition to the 3-day cohort, a separate cohort of 29 patients underwent monitoring with the 7-day WebCardio patch at PSG Hospital during the same duration. The same application, pairing, data transmission, and analysis procedures were followed. Because the monitoring durations differed, the 3-day and 7-day cohorts were analysed separately, and the data were not pooled. For the 7-day cohort, results are reported separately.

RESULTS

Table A Operational definitions of Arrhythmias included in the study

Types of Arrhythmias

Operational definitionᵃ

Atrial fibrillation (AF)

Irregularly irregular rhythm with absence of discrete, consistent P waves, sustained for ≥75 seconds (the minimum AF duration specified in the device's recording settings)

Supraventricular tachycardia (SVT)

≥3 consecutive supraventricular beats at a rate ≥75 beats/min

Non-sustained ventricular tachycardia (NSVT)

≥3 consecutive wide QRS complexes of ventricular origin at a rate ≥120 beats/min; all episodes in this cohort were brief (3–8 beats, approximately 1–3 seconds) and none met criteria for sustained VT

Pause

Sinus pause or atrioventricular block-related pause ≥2.5 seconds

Atrioventricular (AV) block

second-degree (Mobitz type I or II), or third-degree (complete) atrioventricular block, per standard electrocardiographic criteria

ᵃThresholds were confirmed from individual monitoring reports

Clinical characteristics of the study population:

A total of 1076 consecutive patients who presented to the outpatient department with symptoms suggestive of paroxysmal arrhythmias were enrolled. The mean age of the study population was 55.86 ± 19.60 years. Most patients were older than 60 years (553; 54.16%), followed by those aged 40–59 years (296; 28.99%) and 18–39 years (172; 16.85%). There was a male predominance, with 639 male patients (59.39%) and 437 female patients (40.61%).In the separate 7-day cohort (n = 29; mean age 66.41 ± 14.71 years; 16 male [55.17%] and 13 females [44.83%]).  The clinical characteristics of the study population are summarized in Table 1 and Table 4 respectively.

 

Arrhythmias detected:

The recordings were of good quality with very little artifact, as shown in Fig. 2. Analysis of WebCardio patch recordings identified a total of 1076procedures,  As shown in Table 2, supraventricular tachycardia (SVT) was the most common arrhythmia detected (472 events), followed by ventricular tachycardia (VT; 191 events), second-degree atrioventricular block (AVB; 109 events), pauses (100 events), atrial fibrillation (AF; 89 events), complete heart block (CHB; 37 events), and >2:1 AVB (4 events). A substantial proportion of arrhythmias were first detected after 24 hours and would therefore likely have been missed by a routine 24-hour Holter recording. As shown in Table 3, 30.95% of AF, 35.17% of SVT, 40.84% of pauses, and 44.0% of VT events were detected after the first day. Several of these findings have direct clinical relevance: CHB, higher-grade AV block, and pauses are conventionally evaluated for pacing, VT warrants assessment for underlying structural heart disease and arrhythmic risk, and AF informs stroke-prevention decisions. These values are event counts; the number of individual patients affected by each arrhythmia was not derived.

Table 1. Clinical characteristics of the study population

Variable

Overall (N = 1076)

Percentage (%)

Age (years), mean ± SD

(Mean SD) 55.86 ± 19.60

18-39

172

16.85

40-59

296

28.99

>-60

553

54.16

Male, n (%)

639

59.39

Female, n (%)

437

40.61

Note: Values are presented as mean ± standard deviation or n (%). The age-group percentages are calculated using the number of patients with available age-group data as the denominator.

 

Table 2. Distribution of arrhythmias detected during WebCardio patch monitoring

Total monitored

1076

AFib

89

Pauses

100

VT

191

SVT

472

2° AVB Type

109

>2:1 AVB

4

CHB

37

         Note: Values indicate the number of arrhythmic events detected. AFib, atrial fibrillation; AVB, atrioventricular block; CHB, complete heart block; SVT, supraventricular tachycardia; VT, ventricular tachycardia.

 

Table 3. Proportion of arrhythmic events detected after the first 24 hours of monitoring

Type

% Cases Detected After 24 hrs

AF

30.95%

SVT

35.17%

Pause

40.84%

VT

44.0%

      Note: Values represent the percentage of each arrhythmia type first detected after 24 hours of continuous WebCardio patch monitoring. AF, atrial fibrillation; SVT, supraventricular tachycardia; VT, ventricular tachycardia.

 

7-day monitoring cohort: A separate cohort of 29 patients underwent 7-day WebCardio patch monitoring and is reported independently of the 3-day cohort. The mean age was 66.41 ± 14.71 years (range 17–87 years), and most patients were aged 60 years or older (22; 78.57%). There were 16 male patients (55.17%) and 13 female patients (44.83%) (Table 4). The most frequent indications were presyncope (6; 20.69%), stroke (4; 13.79%), syncope (3; 10.34%), and chest pain (3; 10.34%). At least one arrhythmia was detected in 25 of 29 patients (86.21%), while no arrhythmia was detected in 4 patients (13.79%). As shown in Table 5, SVT was the most common finding (19 patients; 65.52%), followed by AV block (9; 31.03%), pauses (9; 31.03%), VT (4; 13.79%), junctional beats (3; 10.34%), and AF (1; 3.45%). Among patients monitored for syncope (n = 3), pauses and AV block were each recorded in 2; among those with presyncope (n = 6), pauses and AV block were each recorded in 2 and VT in 1; and VT was recorded in 1 of 2 patients with palpitation. These findings provide the relevance of monitoring for longer duration to detect those clinically significant arrhythmias which would have been missed by traditional holter.

 

Table 4. Clinical characteristics of the 7-day monitoring cohort

Variable

7-day (N = 29)

Percentage (%)

Age (years), mean ± SD

66.41 ± 14.71

Age range (years)

18–87

18-39

2

3.57

40-59

5

17.86

>=60

22

78.57

Male, n (%)

16

55.17

Female, n (%)

13

44.83

Note: Values are presented as mean ± standard deviation or n (%). Age-group percentages are calculated using the 29 patients with available age-group data

 

Table 5. Arrhythmias detected during 7-day WebCardio patch monitoring (patients, N = 29)

Arrhythmia

Patients, n

%

Any arrhythmia

25

86.21

SVT

19

65.52

AV block (AVB)

9

31.03

Pause

9

31.03

First-degree AV block

8

27.59

VT

4

13.79

Junctional beats

3

10.34

Junctional rhythm

1

3.45

AF

1

3.45

No arrhythmia

4

13.79

Note patients could have more than one arrhythmia type, so categories overlap. Percentages use N = 29. AF, atrial fibrillation; AVB, atrioventricular block; SVT, supraventricular tachycardia; VT, ventricular tachycardia.

 

Figure 4 First detection all arrhythmias Age wise distribution of detected arrythmias and first day detection (3-Day Monitoring)

DISCUSSION

Several devices are available in the market today for evaluation of paroxysmal arrhythmias in ambulatory patients1. Traditionally, Holter recording is done for 24 hours. Continuous ECG recording is useful, but it is cumbersome because multiple leads are attached to the patient, and it is often affected by artifacts and reduced patient compliance. Event recorders are another type of device available, but they must be worn continuously and do not provide continuous ECG data2,3. The patient must trigger the device during the event, which may not always be possible, and these devices may not capture silent asymptomatic arrhythmias. Loop recorders also provide only a fixed length of recording data before and after a predefined arrhythmia2,3. Mobile cardiac telemetry devices can record continuous ECG for 30 days, but they are costly and bulkier, limiting their use. Implantable loop recorders are invasive and costly and therefore cannot be used frequently. External patch monitors like Zio Patch are available in the market and are easy to use and wearable without disconnection for prolonged periods4, but they are not widely available in our setting. A similar device, the WebCardio patch monitoring system, is waterproof and easily attached to the chest wall without dermal irritation5,6. It can be removed during bathing and reattached by the patient during the monitoring period and can be used for monitoring up to 7 days. In our study, we evaluated the effectiveness of patch monitoring in detecting arrhythmias in patients with infrequent paroxysmal arrhythmic symptoms. WebCardio is a cloud-based cardiac rhythm monitoring device that can be easily applied to the patient’s chest wall, with good compliance and acceptance for prolonged monitoring. In this study, 1076 arrhythmic events were recorded, with SVT being the most common event (472), followed by VT (191), second-degree AV block (109), pauses (100), AFib (89), complete heart block (37), and >2:1 AV block (4). Notably, a considerable proportion of arrhythmias were first detected after the initial 24 hours: AF 30.95%, SVT 35.17%, pauses 40.84%, and VT 44.0%. These findings suggest that traditional 24-hour Holter recording may miss a clinically meaningful proportion of arrhythmias, supporting the potential role of convenient, wearable patch monitoring in the evaluation of paroxysmal arrhythmias, although a direct comparison with Holter monitoring was not performed in this study.Long-term, continuous, uninterrupted cardiac recording may improve arrhythmia detection compared with current standard recording technologies. Continuous ECG recording for 24–48 hours with multi-lead Holter monitors is commonly used as an initial option for screening and detecting cardiac arrhythmias7,8. Although multi-lead ECG monitors may offer advantages over single-lead monitors in selected situations, such as distinguishing ventricular tachycardia from aberrant atrial rhythms, limitations related to patient comfort, adherence, and signal quality may reduce arrhythmia detection within a 1–2-day recording window and may require repeat monitoring. Other longer-duration external monitoring technologies, such as external loop recorders and mobile cardiovascular telemetry, may improve detection compared with Holter monitoring because of longer recording periods. However, their use may be limited by the need to recharge or replace batteries, replace lead-wire patches, or remove the device during activities such as showering or physical activity9. Finally, although implantable cardiac monitors have shown promise for long-term arrhythmia detection, their invasive nature and high cost may limit widespread use and make them unsuitable as a first-line monitoring option for many patients.

 

In the separate 7-day cohort of 29 patients (mean age 66.41 ± 14.71 years), at least one arrhythmia was detected in 86.21%, most frequently SVT (65.52%), AV block (31.03%), pauses (31.03%), and VT (13.79%). These findings indicate that prolonged patch monitoring can capture a broad spectrum of arrhythmias, including potentially clinically significant bradyarrhythmias and ventricular tachycardia which would be missed by traditional holter. The 3-day and 7-day cohorts were analysed independently and were not pooled or directly compared. in addition,the 7-day cohort was small, so larger comparative studies are needed to define the incremental yield of 7-day over 3-day monitoring.

 

Clinical implications: The arrhythmias identified through extended patch monitoring have direct management relevance and illustrate the clinical utility of prolonged rhythm surveillance. Documentation of pauses, higher-grade atrioventricular block, or complete heart block in patients with syncope or presyncope can establish a bradyarrhythmic cause of symptoms and guide appropriate evaluation for permanent pacing. Detection of ventricular tachycardia enables risk stratification and may prompt assessment for underlying structural heart disease, initiation or adjustment of antiarrhythmic therapy, or referral for electrophysiological evaluation.

CONCLUSION

Prolonged continuous patch-based ECG monitoring with the WebCardio system demonstrated clinical utility by detecting a substantial burden of clinically actionable arrhythmias, including SVT, AV block, pauses, VT, and AF. The finding that many events were first identified after 24 hours indicates that shorter recordings may under-detect intermittent arrhythmias, while the high arrhythmia-detection rate in the 7-day cohort further supports the value of extended monitoring. Detection of conduction disease, pauses, VT, and AF provides clinically meaningful information that can inform pacing evaluation, anticoagulation, antiarrhythmic therapy, and electrophysiological referral. Although downstream management decisions and outcomes were not assessed, these results suggest that WebCardio patch monitoring can provide actionable diagnostic information in patients with suspected paroxysmal arrhythmias; larger prospective multicentre studies are warranted to confirm these findings and define the incremental yield of 7-day over 3-day monitoring.

REFERENCES

1.      Yenikomshian M, Jarvis J, Patton C, Yee C, Mortimer R, Birnbaum H, Topash M. Cardiac arrhythmia detection outcomes among patients monitored with the Zio patch system: a systematic literature review. Curr Med Res Opin. 2019 Oct;35(10):1659-1670. doi: 10.1080/03007995.2019.1610370. Epub 2019 May 28. PMID: 31045463.

2.      Sivakumaran S, Krahn AD, Klein GJ, et al. A prospective randomized comparison of loop recorders versus Holter monitors in patients with syncope or presyncope. Am J Med. 2003;115:1–5.

3.      Lobodzinski SS, Laks MM. New devices for very long-term ECG monitoring. Cardiol J. 2012;19:210–214.

4.      Barrett PM, Komatireddy R, Haaser S, et al. Comparison of 24-hour Holter monitoring with 14-day novel adhesive patch electrocardiographic monitoring. Am J Med. 2014;127:95.e11–e17.

5.      Robinson J, Snyder C. Ambulatory arrhythmia detection with Zio XT Patch in pediatric patients. Cardiol Young. 2017;27:S1–S653.

6.      Karunadas CP, Mathew C. Comparison of arrhythmia detection by conventional Holter and a novel ambulatory ECG system using patch and Android App, over 24 h period. Indian Pacing Electrophysiol J. 2020 Mar-Apr;20(2):49-53. doi: 10.1016/j.ipej.2019.12.013. Epub 2019 Dec 19. PMID: 31866554; PMCID: PMC7082686.

7.      Reed MJ, Grubb NR, Lang CC, et al. Diagnostic yield of an ambulatory patch monitor in patients with unexplained syncope after initial evaluation in the emergency department: the PATCH-ED study. Emerg Med J. 2018;35:477–485.

8.      Schreiber D, Sattar A, Drigalla D, et al. Ambulatory cardiac monitoring for discharged emergency department patients with possible cardiac arrhythmias. WestJEM. 2014;15:194–198.

9.      Steinhubl SR, Waalen J, Edwards AM, et al. Effect of a homebased wearable continuous ECG monitoring patch on detection CARDIAC ARRHYTHMIA DETECTION MONITORED WITH PATCH SYSTEM of undiagnosed atrial fibrillation: the mSToPS randomized clinical trial. JAMA. 2018;320:146–155.

Recommended Articles
Research Article
Beyond the Transvenous Lead: Real-World Experience with Leadless Pacing in Five Patients
...
Published: 06/11/2026
Download PDF
Research Article
Diagnostic Utility of Right Ventricular Free-Wall Longitudinal Strain by Two-Dimensional Speckle-Tracking Echocardiography for Identifying Proximal Right Coronary Artery Lesions in Acute Inferior Myocardial Infarction
...
Published: 29/10/2026
Download PDF
Research Article
Comparison of in vitro activity of cefepime-enmetazobactam and other carbapenem-sparing agents for ESBL producing gram-negative bacteria
...
Published: 02/11/2026
Download PDF
Original Article
Clinical Profile and Predictors of Severe Dengue in Adult Patients at a Tertiary Care Teaching Hospital: A Hospital-Based Observational Study
Published: 28/03/2023
Download PDF
Chat on WhatsApp
Copyright © EJCM Publisher. All Rights Reserved.